ELAN 30s

Last update: 15.12.2021, 09:56:10
Fully Automated ELISA Microstrip Processor
With the fully automated ELAN 30s, you have all the automation you need to achieve greater workflow efficiency.
 
Complete Solution for Your ELISA Workflow
Every laboratory is different. The ELAN 30s allows users to develop new, flexible analytical workflows by taking advantage of smart scheduling software. By combining up to 6 different assays in a single run users can break the traditional 2 or 4 plate mold and create ‘clinical panel’ based workflows. Multiple reporting formats enhance the final steps of the process and allow the user to report faster with more convenience.
 
DIMENSIONS
  • Length: 754 mm
  • Breadth: 520 mm
  • Height: 888 mm (dome open)
  • Weight: 65 kgs
  • Shipping weight: 95 kgs
     
POWER SUPPLY
  • Voltage: AC 110 - 240 V
  • Frequency: 50 hZ / 60 hZ
  • Power consumption: 300 W
GENERAL SPECIFICATIONS
  • Self-check at startup: Yes
  • Assays per batch: Up to 6 tests
  • Number of strips: 30 × 8 = 240 microwells
  • Sample tube size: 12-13 mm × 75 mm (5ml)
  • Reagent bottle: 24 × 20 ml
  • Controls & Calibrators: 30 × 1.5 ml
  • Location: 41 to 58 can be used for loading reagents in 5mL test tubes
  • Sample dilution: In test tubes
  • Tip capacity: 204 tips
READER SPECIFICATIONS
  • Dynamic range: 0.0 - 3.0 O.D
  • Wavelength range: 400 - 800 nm
  • Filter slots: 405 nm, 450 nm, 492 nm, 630 nm & 3 open positions
  • Reading mode: Single & Dual
  • Reading time: 20 sec per strip for single wavelength
  • Precision: < 1% C.V at 0.5 to 2.0 O.D
  • Accuracy: ± 1 or 2.5% (0.000 to 3.00) whichever is greater
WASHER SPECIFICATIONS
  • Manifold confi guration: 8 for aspiration & 8 for dispensing
  • Wash cycle: 1 to 9 (repeatable)
  • Residual volume: < 6 μL
  • Wash buffer capacity: 1 × 4L & 2 × 1L
  • DW Can: 1 X 20L
  • Low buffer alarm: Yes
  • Soak time: 20 to 60 sec
  • Dispense pressure: Pre-set
  • Rinse function: Yes
  • Waste Can: 1 × 20L
INCUBATOR SPECIFICATIONS
  • Temperature range: 37°C or R.T
  • Temperature accuracy: ± 1°C across plate @37°C
  • Shaker facility: Yes
  • Incubator time: Programmable
  • Time to reach temperature: < 5 min
  • Temperature monitoring: Yes
PIPETTING SPECIFICATIONS
  • Dispensing mode: Fixed needle or disposable tips
  • Sample tips range: 300 μL (5-200 μL)
  • Reagent tip range: 300 μL (10-200 μL)
  • Sample tip precision: < 3% CV at any operating volume above 10 μL
  • < 2% CV at 100 μL
  • Reagent tip precision:
  • < 1% CV at 100 μL
PROCESS SECURITY
  • Liquid level sensor:Yes
  • Level sensor system: Conductive
  • Clot detection: Yes
  • Tip detection: Yes
  • Alarms: Yes
SOFTWARE
  • Controlling software: ELAN 30s
  • Work protocols (assay): Unlimited
  • Data Processing: Quantitative & Qualitative
  • Levey-Jennings: Yes
  • Westgard rules:Yes
  • Process reporting: Event log + error log
  • Automatic error recovery: Yes
  • Password access control: Yes
CONSUMABLES
  • Probe clean solution : XL Wash
  • Carbonated disposable tips: Sample tips, Reagent tips
 

Photo gallery